Qualifying* small and medium-sized US manufacturers, that received IAC or Combined Heat and Power Technical Assistance Partnership assessments between 2018 and 2024, can now apply for grants for implementation of assessment recommendations up 50% of qualifying* project costs with a maximum of $300,000 per manufacturer.
IMPORTANT: The grant program now operates on a rolling basis and applications may be submitted at any time through the year, with quarterly reviews.
*See Full Grant Solicitation for full qualification and selection details.
IAC Center | University of Miami |
---|---|
Assessment Year | 2014 |
Principal Product | Biotechnology |
NAICS | 334516 - Analytical Laboratory Instrument Manufacturing |
SIC | 3826 - Analytical Instruments |
Sales | $10,000,000-$50,000,000 |
# of Employees | 80 |
Plant Area (Sq.Ft.) | 40,000 |
Annual Production | 500,000 Units |
Production Hrs. Annual | 2,600 |
Location (State) | PR |
Source | Yearly Cost | Usage | Unit | Unit Cost |
---|---|---|---|---|
Electricity Usage | $942,255 | 4,363,841 | kWh | $0.22 |
Electricity Demand | $65,350 | 8,131 | kW-months/year | $8.04 |
Electricity Fees | $82,644 | - | $ | - |
TOTAL ENERGY COSTS | $1,090,249 | |||
RECOMMENDED SAVINGS* | $180,941 | *Non-energy impacts included in savings. See recommendations below |
||
IMPLEMENTED SAVINGS* | $150,194 |
#|Description |
Savings Cost |
Status | Electricity Usage | Electricity Demand | Electricity Fees | Electricity Usage | Electricity Demand | Electricity Fees | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
$ | kWh | $ | kW-months/year | $ | $ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
01: 2.7425 CLEAN OR COLOR ROOF TO REDUCE SOLAR LOAD |
Savings: $66,980 Cost: $10,999 |
$57,162 | 264,637 | $9,818 | 1,221 | - | - |
$57,162 264,637 kWh |
$9,818 1,221 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
02: 2.7221 LOWER TEMPERATURE DURING THE WINTER SEASON AND VICE-VERSA |
Savings: $45,927 Cost: $80 |
$39,195 | 181,459 | $6,732 | 838 | - | - |
$39,195 181,459 kWh |
$6,732 838 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
03: 2.6218 TURN OFF EQUIPMENT WHEN NOT IN USE |
Savings: $17,716 Cost: $3,100 |
$15,119 | 69,997 | $2,597 | 323 | - | - |
$15,119 69,997 kWh |
$2,597 323 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
04: 2.4146 USE ADJUSTABLE FREQUENCY DRIVE OR MULTIPLE SPEED MOTORS ON EXISTING SYSTEM |
Savings: $14,333 Cost: $7,040 |
$12,232 | 56,630 | $2,101 | 261 | - | - |
$12,232 56,630 kWh |
$2,101 261 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
05: 2.7447 INSTALL VINYL STRIP / HIGH SPEED / AIR CURTAIN DOORS |
Savings: $11,159 Cost: $2,206 |
$10,276 | 47,771 | $883 | 110 | - | - |
$10,276 47,771 kWh |
$883 110 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
06: 2.3212 OPTIMIZE PLANT POWER FACTOR |
Savings: $5,083 Cost: $7,680 |
- | - | $5,083 | 632 | - | - |
- - kWh |
$5,083 632 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
07: 2.7142 UTILIZE HIGHER EFFICIENCY LAMPS AND/OR BALLASTS |
Savings: $4,680 Cost: $2,160 |
$4,246 | 19,656 | $434 | 54 | - | - |
$4,246 19,656 kWh |
$434 54 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
08: 2.7142 UTILIZE HIGHER EFFICIENCY LAMPS AND/OR BALLASTS |
Savings: $3,949 Cost: $13,170 |
$3,370 | 15,600 | $579 | 72 | - | - |
$3,370 15,600 kWh |
$579 72 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
09: 2.7496 INSTALL PARTITIONS TO REDUCE SIZE OF CONDITIONED SPACE |
Savings: $3,341 Cost: $600 |
$3,179 | 14,716 | $162 | 20 | - | - |
$3,179 14,716 kWh |
$162 20 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
10: 2.7135 INSTALL OCCUPANCY SENSORS |
Savings: $3,162 Cost: $800 |
$2,696 | 12,480 | $466 | 58 | - | - |
$2,696 12,480 kWh |
$466 58 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
11: 2.7444 CLOSE HOLES AND OPENINGS IN BUILDING SUCH AS BROKEN WINDOWS |
Savings: $1,914 Cost: $520 |
$1,633 | 7,561 | $281 | 35 | - | - |
$1,633 7,561 kWh |
$281 35 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
12: 4.821 PAY BILLS ON TIME TO AVOID LATE FEES |
Savings: $1,523 Cost: $240 |
- | - | - | - | $1,523 | - |
- - kWh |
- - kW-months/year |
$1,523 - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
13: 2.4236 ELIMINATE LEAKS IN INERT GAS AND COMPRESSED AIR LINES/ VALVES |
Savings: $727 Cost: $1,660 |
$443 | 2,049 | $284 | 35 | - | - |
$443 2,049 kWh |
$284 35 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
14: 2.4231 REDUCE THE PRESSURE OF COMPRESSED AIR TO THE MINIMUM REQUIRED |
Savings: $447 Cost: $80 |
$353 | 1,634 | $94 | 12 | - | - |
$353 1,634 kWh |
$94 12 kW-months/year |
- - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
TOTAL RECOMMENDED |
Savings: $180,941 Cost: $50,335 |
$149,904 | 694,190 | $29,514 | 3,671 | $1,523 | - | $149,904 694,190 kWh |
$29,514 3,671 kW-months/year |
$1,523 - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
TOTAL IMPLEMENTED |
Savings: $150,194 Cost: $39,969 |
$122,584 | 567,512 | $26,087 | 3,245 | $1,523 | - | $122,584 567,512 kWh |
$26,087 3,245 kW-months/year |
$1,523 - $ |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Savings Cost |
$ | kWh | $ | kW-months/year | $ | $ | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Electricity Usage | Electricity Demand | Electricity Fees | Electricity Usage | Electricity Demand | Electricity Fees |